Swiss Measuring Instruments

Setting Bench Alesta N

Motorized setting bench for the control and calibration of measuring means with fixed and variable dimensions

Motorized Alesta N is the ideal instrument for the control and calibration of measuring means with variable dimensions. Its high precision and finishing level position it as a top-of-the-range model for workshop or laboratory use. It is the perfect answer to current needs in the manufacturing field.

This new model is based on the already well-known Trimos Horizon Setting line, but a new electronics and innovative color display have been incorporated. A modern and ergonomic design, as well as great ease of use, guarantee optimal and high-productivity measuring results.

Trimos proposes a complete range of instruments. All measuring ranges are direct, which means that the whole measuring range is available without adjustment or intermediary re-calibration. A large selection of easily interchangeable accessories completes the possible applications and guarantees a correct positioning of the instruments to be calibrated.

Perfect for setting and zeroing gauges in the shopfloor, they can also be used as a calibration machine in compliance with ISO standards. The adapters ensure objective and precise setting of gages in just a few seconds.

  • Ideal for use in the workshop area
  • Very simple manipulation
  • Motorized carriage movement
  • For inside, outside, 2 and 3 points instruments
  • No need to buy and recalibrate gage blocks and rings
  • Inhouse calibration of instruments
  • Time for setting operation reduced by 90%
  • Connectivity and interface via USB and Ethernet
  • Adapters of ALESTA version compatible

Technical details

ALESTA N3006001000
Application range (external measurements)*mm (in)300 (12)600 (23)1000 (40)
Application range (internal measurements)*mm (in)40 ÷ 340 (1.6 ÷ 13.5)40 ÷ 640 (1.6 ÷ 23)40 ÷ 1040 (1.6 ÷ 41.5)
Max. permissible errors 1)µm1.5 + L(mm)/500
Repeatability, RMPE (2s) 1)µm0.8
Resolutionsmm (in)0.0005 (.0000197)
InterfacesETHERNET and USB
Operational temperature°C10 ÷ +35
Temperature of storage°C-10 ÷ +40
Weightkg80140360

* The measuring range depends on the anvils used
1) Values valid at temperature of 20±0.5 °C and relative humidity of 50±5%.